Skip to main content

Public Health

  • Webpage

    BOD seed

    Seed preparation and addition De-ionized water kills - Adding a seed capsule to deionized water causes an osmotic imbalance that will kill seed organisms. The purified water —containing no salts— rushes across the bacterial cell walls. The bacteria…

  • Webpage

    Sampling and sample handling for BOD

    Water and wastewater samples are biologically active, and the BOD will quickly deteriorate. Minimize the time between sampling and analysis to obtain the best BOD results. Take the following precautions when sampling and handling BOD samples.…

  • Webpage

    BOD resources

    Websites Field Training Manual for Laboratory Analysts: Chapter 7 - BOD Biochemical Oxygen Demand Source: http://home.windstream.net/mikeric/Chap1to9/Chap7%20Biochemical%20Oxygen%20Demand.htm-->

  • Webpage

    BOD quality control

    Limit of detection (LOD) BOD detection limits are theoretically based. A key assumption which the concept of detection for BOD is based is that the LEAST amount of depletion allowable is 2 mg/L. The LOD is also based on the highest volume of sample…

  • Webpage

    BOD sample pre-treatment

    Because the BOD test is a bioassay, it is critical to maintain optimal conditions for the "bugs" (bacteria and other microorganisms) to remain viable. Temperature, pH, oxygen levels and the presence of disinfectants can all influence the outcome of…

  • Webpage

    cBOD

    Nitrogenous Oxygen Demand (NOD) cBOD v. BOD - If you have Nitrogenous Oxygen Demand (NOD) you should consider analyzing cBOD vs. BOD

  • Webpage

    Method Details

    The following are the critical steps in preparing and analyzing samples for BOD testing. The BOD test is a method-defined (empirical) technique. Consequently, the method must be followed carefully, and must be performed consistently from day to day…

  • Webpage

    BOD equipment and supplies

    BOD Incubator Thermostatically controlled at 20± 1°C.

  • Webpage

    Oxygen measurement techniques

    DO Probe Electrochemical method: Composed of two metal electrodes in contact with supporting electrolyte and separated from the test solution by a gas permeable membrane. A constant voltage is placed across the cathode and anode. Oxygen diffuses…

  • Webpage

    BOD dilution water

    BOD dilution water nutrient solutions can be a source of contamination. If you prepare your own solution, make sure you store the phosphate buffer in a refrigerator. Discard any solution if it becomes cloudy or you observe any "chunks" floating in…

Subscribe to Public Health